The George Washington University
School of Engineering and Applied Science
Department of Electrical and Computer Engineering
ECE 11 - Fall 2003
Experiment # 6
Instrument Input Resistance, Analog Oscilloscope Measurements,
Function Generator Specifications, Meter Responses and
Waveforms
Equipment:
1 Keithley Model 175 Digital Multimeter
(DMM)
2 Pair of Banana to mini-grabber
test leads
1 Prototype Bread Board
1 BNC T Connector
1 Banana to BNC Adapter Plug
1 50 Ohm Load
Components:
RL 50 Ohm
R1 1.5 MOhm
Objectives:
-
Determine input resistance of Keithley
Model 175
-
Determine input resistance of Keithley
Model 178
-
Determine input resistance of Tektronix
Model 2213
-
Verify maximum/minimum input voltage
of HP8116A
-
Use Tektronix Model 2213 to measure
AC Voltage
-
Determine the peak voltage values (Vp)
for different periodic waveforms
-
Determine the rms voltage values (Vrms)
for different periodic waveforms
-
Determine the average voltage values
(Vavg) for different periodic waveforms
-
Compare and explain the results obtained
from DMM and analog oscilloscope for measurement of different periodic
waveforms
1. - Input Resistance Measurement
on Keithley Model 175/178
(PreLab) A. Design
-
Design a series circuit that has a voltage source (Vs) and two resistors
(R1 and RM).
-
RM represent input resistance of Keithley Model 175 / 178.
-
Draw this test circuit and annotate it as Figure #1.
-
Derive a general equation in terms of RM assuming the values of Vs, VM
and R1 are known.
-
Label this equation - "Equation 1 - Equation for finding RM"
B. Verification
-
Construct the circuit of Figure #1. Use Vs = 15Vdc and R1=1.5 M Ohm.
-
Use Voltmeter function of Keithley Model 175 to measure VM.
-
Measure Vs, R1 and VM as accurate as you could.
-
Use Equation 1 to find input resistance of Keithley 175 using measured
values Vs, R1 and VM.
-
Repeat step 1 to 4 using voltmeter function of Keithley 178.
-
Find input resistance of Keithley 178.
2. - Input Resistance Measurement
of Tektronix Model 2213
(PreLab) A. Design
-
Design a series circuit that has a voltage source (Vs) and two resistors
(R1 and Rscope).
-
Rscope represent the input resistance of Tektronix 2213.
-
Draw this test circuit and annotate it as Figure #2.
-
Derive a general equation in terms of Rscope assuming the values of Vs,Vscope
and R1 are known.
-
Label this equation - "Equation 2 - Equation for finding Rscope"
B. Verification
-
Construct the circuit of Figure #2. Use Vs = 15Vdc.
-
Use Voltmeter function of Keithley Model 175 to measure Vs, and VR1.
-
Use Tektronix 2213 to measure DC voltage across Rcope.
-
Measure Vs, VR1 and Vscope as accurate as you could.
-
Use Equation 2 to find Rscope.
3. - AC Voltage Measurement
The concept of AC signal is introduced in this section. Function
generator (HP8116A) is used to produce AC voltage. Three most common
AC signals are sine wave, rectangular wave and triangular wave. Oscilloscope
is used to measure AC signals and in this lab analog oscilloscope
(Tektronix 2213) is used.
A. Setting up HP8116A
-
Read HP8116A manuals throughly. The manual can be read HERE.
-
Set up HP 8116A to the folowing specifications.
POWER 1(ON)
MODE
NORM
WAVEFORM depress a button that has sine wave above it.
FREQ
use the range and vernier switch to select 10 kHz
DTY
50%
AMP
16.0 V
OFS
0V
-
Connect the output of function generator to one end of a BNC Cable.
-
Depress the DISABLE button on the function generator panel.
B. Setting up Tektronix 2213 and AC Signal Measurement
-
Read Tektronix 2213 manual throughly. The manual can be read HERE.
-
Read in detailed the section on how to measure AC Voltage.
-
Connect the the other end of BNC cable coming from the function generator
to CH1 input of the oscilloscope.
-
If needed, adjust the trigger control to obtain a stable waveform.
-
Adjust CH1 VOLTS/DIV to display at least one complete sinewave of the signal
on the screen.
-
Adjust SEC/DIV to display at least one complete cycle sinewave of the signal
on the screen.
-
Measure and record the peak to peak voltage value of the waveform shown
on the screen.
-
On the function generator, change AMP = 10.0 mV
-
Repeat step 4-7 above.
-
Put the measurement result in Data Table 3B - AC Voltage Measurement
using Tektronix 2213.
C. Measurement of AC Signal with 50 Ohm Load
-
Draw the function generator (HP8116A) with its output resistance(Rs) connected
to the load (RL). Output resistance of function
generator can be found from the online-manual.
Set the load RL to be 50 Ohm.
-
Label this circuit Figure #3.
-
Derive a general equation to describe the peak to peak voltage across RL.
Label this Equation 3C.
-
Change AMP = 16.0 V on the function generator.
-
Construct Figure #3 and connect the input of function generator to CH1
of oscilloscope through a BNC cable.
-
Measure the peak to peak output voltage across RL.
-
Put the measurement result in Data Table 3B.
4. Meter Responses and Waveforms
(PreLab)
A. Calculation of Vpeak peak, Vrms and Vavg for different periodic waveforms
Show the formula on how to find
the V peak to peak, Vrms and Vavg values for:
1. Sine wave
2. Triangle wave
3. Square Wave
Sketch the general waveform for each wave.
[Hint] Read textbook Chapter 5-6.
Given Vamplitude = 5.0 Vpeak and f = 1 KHz, find nominal values for
those three waveforms. Put all the calculated data in Data Table
#4A.
|
Vpp |
Vp |
Vrms |
Vavg |
| Sine Wave |
|
|
|
|
| Triangle Wave |
|
|
|
|
| square wave |
|
|
|
|
Data Table #4A
B. - Measurement of Vpeak, Vrms
and Vavg for different periodic waveforms
In this section you are asked to
measure Vpp, Vp, Vrms and Vavg of periodic waveforms using the combination
of three instrument: Keithley 175, Keithley 178 and analog oscilloscope
(Tektronix 2213). Read the
instruments
manual on AC Voltage measurement to have a better understanding of
their measurement capabilities.
-
Set up HP 8116A to the folowing specifications.
POWER
1(ON)
MODE
NORM
WAVEFORM depress
a button that has a SINE wave on it.
FREQ
use the range and vernier switch to select 1 kHz
DTY
50%
AMP
5.0 V
OFS
0V
-
Depress the DISABLE button on the function generator panel.
-
Set Tektronix 2213, Keithley Model 175 and Keithley Model 178 in
a way so that the combination of those equipment can measure Vpp, Vp, Vrms
and Vavg output of the function generator ALL AT THE SAME TIME. Put
the measured data in Data Table 4B.
-
Change the waveform setting of the function generator to TRIANGULAR
WAVE. Repeat step 3.
-
Change the waveform setting of the function generator to SQUARE WAVE.
Repeat step 3.
-
Analyze the data of Data Table 2 and answer these questions:
-
Which instrument measures the average value of an input waveform but is
calibrated to read the RMS value of that waveform. (Hint: Check the manual
of each instrument on AC measurement to find this information.)
-
Identify which instrument is best suited to measure rms value at this frequency.
5. Conclusions
-
Compare the measured input resistances
for Keithley 175, Keithley 178 and Tektronix 2213 to the published input
resistances obtained from the instrument's manuals.
-
Is it important for us to know the values
of the instrument's input resistances that we are using in the lab?
Explain.
-
Compare Data Table 4A and Data Table
4B. Explain any differences.
Converted to HTML format by
Faisal Mohd Yasin, cepus@seas.gwu.edu, June and October 2001.
Modified by Yang Cao, ycao@gwu.edu,
Sept. 2003